Clamp-on Flow Meter
FD-R series
SpecsClamp-on Flow MeterFD-R series
Model |
FD-R50 |
FD-R80 |
FD-R125 |
FD-R200 |
|||
Image |
|
|
|
|
|||
Supported pipe diameter |
1又1/2”(40 A), ø44 to ø551.73" to 2.17" |
2 1/2" (65 A), ø64 to ø832.52" to 3.27" |
4" (100 A), ø100 to ø1273.94" to 5.00" |
6" (150 A), ø152 to ø1915.98" to 7.52" |
|||
Supported pipe materials |
Metal / resin* 1 |
||||||
Supported fluids |
Various liquids (i.e. water, oils, chemicals)* 1 |
||||||
Supported fluid temperature |
- 20 to + 120°C- 4.0 to + 248 °F(no freezing on the pipe surface)* 2 |
||||||
Maximum rated flow |
Rated flow velocity range: 5.0 m/s |
Rated flow velocity range: 5.0 m/s |
Rated flow velocity range: 5.0 m/s |
Rated flow velocity range: 5.0 m/s |
|||
Zero cut flow rate |
Zero cut: 0.3 m/s (default)* 3 |
Zero cut: 0.3 m/s (default)* 3 |
Zero cut: 0.3 m/s (default)* 3 |
Zero cut: 0.3 m/s (default)* 3 |
|||
Display method |
Dual row, 5-digit display with white, 14-segment LED; Large status indicator; Output indicators; Stability indicator; Unit indicator |
||||||
Display update cycle |
大约3赫兹 |
||||||
Display resolution |
0.1 / 1 (L / min) |
1 (L / min) |
|||||
Response time |
0.5 s / 1.0 s / 2.5 s / 5.0 s / 10.0 s / 30.0 s / 60.0 s / 120.0 s / 200.0 s (variable) |
||||||
Measurement accuracy |
Between 20 and 100% of F.S. : ±2.0% of RD*4*5 |
||||||
Protection circuit |
权力supply reverse connection protection, Power supply surge protection, Short-circuit protection for each output, Surge protection for each output |
||||||
Zero point error |
±0.5% of F.S.*4*6 |
||||||
Hysteresis |
Variable |
||||||
Flow units |
L / min, m3/ h, gal / min |
||||||
Integrated flow unit display |
1 / 10 / 100 / 1000 / 10000 (L) |
||||||
Pipe temperature measurement accuracy |
±3°C±5.4°F(liquid temperature of - 20 to + 50°C,- 4 to + 122°F) |
||||||
Wiring specifications |
权力supply |
DC power supply: M12 4-pin connector / AC power supply: M4 screw terminal block (selectable) |
|||||
I/O |
When using a DC power supply: M12 4-pin connector / when using an AC power supply: M3 screw terminal block |
||||||
Input/Output (Selectable) |
Output (ch.1/ch.2) |
Control output / Integrated pulse output / Error output / Temperature alarm, NPN / PNP setting switchable,open collector output 30 VDC or less, max. 100 mA / ch., residual voltage: 2.5 V or less*7 |
|||||
Analog output (ch.1/ch.2) |
Flow rate analog output / Temperature analog output and 4 - 20 mA / 0 - 20 mA (selectable), load resistance: 500 Ω or less*7 |
||||||
External input (ch.2) |
Integrated flow reset input / Flow rate zero input / Origin adjustment input (selectable), short-circuit current: 1.5 mA or less, input time: 20 ms or more*7 |
||||||
Rating |
权力voltage |
20 to 30 VDC including 10% ripple (P-P), Class 2 / 100 to 240 VAC - 15% or + 10% (50 / 60 Hz) |
|||||
Current consumption |
When using a DC power supply: 200 mA or less (load current excluded), 400 mA or less (load current included) When using an AC power supply: 15 VA or less |
||||||
Environmental resistance |
Enclosure rating |
IP65 / 67 (IEC60259), IP69K (ISO20653), Enclosure Type 4X (NEMA250) |
|||||
Ambient temperature |
- 20 to + 60°C- 4.0 to 140 °F(no freezing)* 2 |
||||||
相对湿度 |
5 to 90%RH (no condensation) |
||||||
Vibration resistance |
10 to 55 Hz, compound amplitude 1.5 mm0.06", XYZ axes 2 hours for each axis |
||||||
Shock resistance |
100 m/s2, 16 ms pulse, XYZ axes, 1000 times for each axis |
||||||
Dimensions |
Main unit size: 218.5 mm × 66.9 mm × 70.7 mm8.60" × 2.63" × 2.78" |
||||||
Material |
Main unit |
Body: aluminum die-casting + coating / PPS, display: reinforced glass, connectors: SUS304-equivalent |
|||||
Unit rear |
Rubber |
||||||
Upper/lower bracket |
SUS304 |
||||||
Weight |
Main unit: Approx. 1.0 kg |
Main unit: Approx. 1.0 kg |
Main unit: Approx. 1.0 kg |
Main unit: Approx. 1.0 kg |
|||
* 1Liquid must allow for the passage of an ultrasonic pulse, as well as not contain large air pockets or excessive bubbles. Detection may be unstable due to the type and status of the pipes. |
|||||||